Contrast Leak on CT Detector Creates Stroke-Mimicking Artifact

AJNR. American journal of neuroradiologyyesterday

Contrast leaked onto CT detector created peripheral dark artifacts mimicking stroke on cranial exams. Confirmed via phantom; radiologists must recognize this scanner-based artifact for quality control and to avoid misdiagnosis.

  • Peripheral dark attenuation defects on cranial CT can be caused by contrast material on the detector panel.
  • Phantom scanning confirmed the artifact source, emphasizing the need for quality control.
  • Recognition of uncommon scanner-based artifacts is essential to prevent misdiagnosis of acute infarction.